Contact Us:Phone: 080 - 42027704 contact@operonbiotech.com

IVD

Biochemistry

Agatha Touch: Semi Auto Bio Chemistry Analyzer

Semi auto Chemistry Analyzer, Open System with Touch Screen and Keypad Operations.3 Temperature Selection,7 interference filters and Low Aspiration volume of 200-3000 ul, which makes Agatha Touch a Full Package for a small laboratory and is approved by TUV & CMD.

  • Touch Screen/ Keypad Operations.- 7 Filters (340nm, 380nm, 405nm, 505nm, 546nm, 578nm, 620nm) + one open position
  • Flow Cell : Permanent quartz flow cell, minimal volume is 32 µl, And disposable flow cell is selectable.Memory: 200 programmable parameters, 15000 test results”
  • Light Source:6V 10W,halogen lamp of more than 2000 hours life time.
  • Spectrometer: Interference filter.
  • Wavelength: 340nm. 380nm. 405nm. 505nm. 546nm. 578nm. 620nm, and one free position
  • CML Precision: ± 2nm
  • FWHM: c 10nm
  • Stray Light: 0.5%(absorbance;’ 2.5)
  • Absorbance Range: -0.214 – 3.000
  • Absorbance Linearity: ± 1%
  • Absorbance Repeatability: ± 0.001
  • Absorbance Resolution: 0.0001
  • Precision: c 1.0%
  • Cany Over: 1%
  • Temperature: 25 °C. 30°C. 37’C, ±0.3 °C.
  • Flow Cell: Permanent quartz flow cell, minimal volume is 32 µl, and disposable flow cell is selectable
  • Sampling System: Pressure auto releasing pumps ensure the accuracy of suction volume
  • Test Method: End point, Fixed time, Kinetic, Factor,Dual wavelengths, Multi-Standards,Bichromatic.
  • Parameter Setting: Patients profile and calculated item can be set easily. Aspiration Volume: 200 µl – 3000 µl.
  • Sample Volume: 10 µl – 50 µl.
  • Reaction Curve: Real time display the reaction curve
  • Memory: Able to store more than 200 parameters while 55 parameters of them are preprogrammed,able to save more than 15000 test results and auto save the results if the power is off during working.
  • QC: Auto draw the QC figure
  • Reagent: Open reagent
  • Display: LCD with high resolution
  • Input: Key button and touch screen
  • Printer: Inner thermal printer,able to be connected to external printer
  • Report Content: Absorbance,concentration,activity,detailed patient information and abnormal results marking.
  • Interface: RS-232 serial port,able to be connected to computer
  • Working Condition: Temperature:10’C -40’C,Relative Humidity: 20% – 80%
  • Voltage: AC 85-264V,auto adaption according to the input voltage
  • Power: ≤ 100W
  • Quality System: Recognized by TUV and CMD.

PA54 Specific Protein Analyzer with TUV,CE ,NGSP Certification

“PA54 Specific Protein Analyzer performs tests in just three simple steps.it helps in monitoring of Inflamation,Rheumatism,Diabetes, Hemostasis, Renal Function,Cardivascular and immunological diseases.”

  • 5.6 inches color touch screen
  • Calibration by magnetic card
  • 4 independent channels
  • Disposable cuvette, no carryover
  • Can be extended more than 30 parameters
  • Support LIS and barcode
  • Less and easy maintenance.
  • Tests/Kit: 25T/2×25T/50T/100T
  • Principle : Nephlometry
  • Throughput : 160 samples/hour
  • Parameters: HS-CRP, CRP, HbA1c, mALB, D-Dimer, Anti-CCP, RF, ASO, CYS-C, IgA, IgM, IgG, C3, C4, etc. Can be extended more than 30 parameters
  • Sample type : Whole blood, serum, plasma, urine etc.
  • Light source: Diode laser, 650nm wavelength
  • Calibration : Magnetic card, make the result more accurately
  • Cuvette : Disposable cuvette
  • Testing channel : 4 independent channels
  • Reaction time : 1.5~3min
  • Language : English, Russian
  • Display : 5.6 inches color touch screen
  • Memory : 100,000 results storage
  • Print : Auto and manual printing, thermal printer
  • Working temperature :10~30°C
  • Precision : CV≤5%
  • Power supply : AC 100-240V 50/60Hz
  • Size : 345mm×352mm×180mm
  • Net weight : 5.4kg
  • Support LIS and barcode
  • NGSP certified

BS-120: Fully Automated Biochemistry Analyzer

” Discrete, random access, fully automated with 100 tests/ hour , up to 300 tests/ hr Up to 28 on-board chemistries and 4 ions, 8 wavelengths: 340 – 670 nm and Bi- directional LIS interface . ”

  • Discrete, random access, fully automated
  • 100 tests per hour, up to 300 tests per hour with ISE
  • Up to 28 onboard chemistries and 4 ions e Refrigerated reagent compartment
  • Automatic probe cleaning, liquid level detection & collision protection
  • 8 wavelengths: 340-670nm
  • Automatic dilution for abnormal sample o External bar code reader (optional) oBi-directional LIS interface
  • Automatic, Discrete, Random Access
  • STAT sample priority
  • ISE Module (optional): Measuring K+, Na+, Cl−
  • Throughput: Constant 200 tests/hour (without ISE), up to 330 tests/hour with ISE
  • Principles: Absorbance photometry, Turbidimetry, Ion Selective Electrode technology
  • Methodology: End-point, Fixed-time, Kinetic, optional ISE Single/Dual reagent chemistries, monochromatic/bichromatic Linear/non-linear multi-point calibration
  • Programming: Open system with user defined profiles and calculation chemistries Original system pack reagents ready to use
  • Reagent/Sample Handling: Reagent/Sample tray: 40 positions for reagents and 40 positions for samples in cooling compartment (2~12˚C)
  • Optical System:
  • Light Source: Halogen-tungsten lamp
  • Photometer: Grating system, reversed optics
  • Wavelength: 12 wavelengths, 340nm, 380nm, 412nm, 450nm, 505nm, 546nm, 570nm, 605nm, 660nm, 700nm, 740nm and 800nm
  • Absorption range: 0~3.3Abs (10mm conversion)
  • Resolution: 0.0001Abs
  • Reaction System:
  • Reaction rotor: Rotating tray, containing 80 cuvettes
  • Cuvette: Reusable, optical length 5mm
  • Reaction volume: 150~500μl
  • Reaction temperature: 37˚C
  • Temperature fluctuation: ±0.1˚C
  • Mixing System: Independent mixing bar
  • Reagent volume:
  • R1: 10~350μl, step by 1 μl
  • R2: 10~200μl, step by 1 μl
  • Sample volume: 2~45μl, step by 0.1 μl
  • Reagent/Sample probe:
  • Liquid level detection and tracing, vertical & horizontal collision protection and inventory checking
  • Probe cleaning: Automatic washing for both interior and exterior Carry-over < 0.1%
  • Automatic sample dilution: Pre-dilution and post-dilution Dilution ratio up to 1: 200
  • Internal Bar Code Reader (optional): Used for sample and reagent programming Applicable to various bar code systems of Codabar, ITF (Interleaved Two of Five), code128, code39, UPC/EAN, Code93 Capable to communicate with LIS in bi-directional mode
  • Cuvette Washing: 8-step washing station with pre-warmed detergent and water
  • Control and Calibration: Calibration mode: Linear (one-point, two-point and multi-point), Logit-Log 4P, Logit-Log 5P, Spline, Exponential, Polynomial, Parabola
  • Control software: Westgard multi-rule, Cumulative sum check, Twin plot
  • Operation Unit:
  • Operation system: Windows® XP Professional/Home SP2 or above Windows® 7
  • Interface: RS-232
  • Working Conditions:
  • Power Supply: AC 200~240V, 50/60Hz, ≤1500VA or AC 100~130V, 50/60Hz, ≤1500VA
  • Temperature: 15-30˚C for operation
  • Humidity: 35-85% RH
  • Water consumption: ≤4.5L/hour
  • Dimension: 860mm (W) x700mm (D) x625mm (H)
  • Weight: 130 Kg

BS-200:Fully automated Biochemistry Analyzer

Discreet random access,200 test/ hour, up to 330 test/hour with ISE (Optional ISE module),refrigerated compartment with bi-directional LIS ,Automatic probe cleaning, liquid level detection, collision protection.

  • Discrete, random access, fully automated, 200 tests per hour, upto 330 tests per hour with ISE Optional for ISE module and internal bad code reader.
  • 40 positions for samples and reagents respectively
  • Automatic probe cleaning, liquid level detection, collision protection.
  • Reserve optic system with 8 wavelengths: 340~670 nm.
  • Refrigerated reagent and sample compartment.
  • Bi-directional LIS Interface.
  • System Function:
  • Automatic, Discrete, Random Access
  • STAT sample priority
  • Throughput: Upto 200 tests/hour (without ISE), up to 330 tests/hour with ISE
  • Measuring Principles: Absorbance photometry, Turbidimetry, Ion Selective Electrode technology
  • Methodology: End-point, Fixed-time, Kinetic, optional ISE Single/Dual reagent chemistries, monochromatic/bichromatic Linear/non-linear multi-point calibration
  • Programming: Open system with user defined profiles and calculation chemistries Original system pack reagents ready to use
  • Reagent/Sample Handling: Reagent/Sample tray: 40 positions for reagents and 40 positions for samples in cooling compartment (4~15˚C)
  • Reagent volume :
  • R1: 180 to 450µl, step by 1 µl
  • R2: 3–450µl, step by 1µl
  • Sample volume : 3~45µl, step by 0.5 µl
  • Reagent/Sample probe: Liquid level detection and tracing, vertical & horizontal collision protection and inventory checking
  • Probe cleaning: Automatic washing for both interior and exterior Carry-over < 0.1%
  • Automatic sample dilution: Pre-dilution and post-dilution Dilution ratio up to 1: 200
  • Internal Bar Code Reader (optional): Used for sample and reagent programming Applicable to various bar code systems of Codabar, ITF (Interleaved Two of Five), code128, code39, UPC/EAN, Code93 Capable to communicate with LIS in bi-directional mode
  • ISE Module (optional): Measuring K+, Na+, Cl−
  • Reaction System:
  • Reaction rotor: Rotating tray, containing 80 cuvettes
  • Cuvette: Reusable, optical length 5mm
  • Reaction volume: 150~500μl
  • Reaction temperature: 37˚C
  • Temperature fluctuation: ±0.1˚C
  • Mixing System: Independent mixing bar
  • Optical System:
  • Light Source: Halogen-tungsten lamp
  • Photometer: Reversed optics, static fiber sport photometry
  • Wavelength: 340nm, 405nm, 450nm, 510nm, 546nm, 578nm, 630nm, 670nm
  • Absorption range: 0~4.0Abs (10mm conversion)
  • Resolution: 0.0001Abs
  • Control and Calibration:
  • Calibration mode: Linear (one-point, two-point and multi-point), Logit-Log 4P, Logit-Log 5P, Spline, Exponential, Polynomial, Parabola
  • Control software: Westgard multi-rule, Cumulative sum check, Twin plot
  • Operation Unit:
  • Operation system: Windows® XP Professional/Home SP2 or above Windows® 7
  • Interface: RS-232
  • Working Conditions:
  • Power Supply: AC 200~240V, 50/60Hz, ≤1000W or AC 100~130V, 50/60Hz, ≤1000W
  • Temperature: 15-30˚C for operation
  • Humidity: 35-80% RH
  • Water consumption: ≤3.5L/hour
  • Dimension: bench top: 34inch (W) x28inch (D) x25inch (H)
  • Weight:
  • Bench top : 116 Kg
  • Cabinet (optional) : 51 kg
BS-200E

BS-200E: Fully automated Biochemistry Analyzer with On-Board laundry

Function: Automatic, Discrete, Random Access STAT sample priority Throughput: Constant 200 tests/hour (without ISE),upto 330 tests/hour with ISE Principles: Absorbance photometry, Turbidimetry, Ion Selective Electrode technology

  • Constant 800 photometric tests per hour, up to 600 tests per hour for ISE (K, Na, Cl).
  • 24-hour refrigeration for reagent tray and inner sample tray.
  • Permanent cuvettes with 8-stage auto-washing station
  • Two 3-head mixing bar units.
  • Minimum 100μl reaction volume.
  • Reversed grating system with 12 wavelengths.
  • Bubble detection for reagent and sample probes.
  • Support up to 4000m altitude application (external air pump optional).
  • Bi-directional LIS interface.”
    • System Function: Automatic, Discrete, Random Access STAT sample priority
    • ISE Module (optional): Measuring K+, Na+, Cl−
    • Throughput: Constant 200 tests/hour (without ISE), up to 330 tests/hour with ISE
    • Principles: Absorbance photometry, Turbidimetry, Ion Selective Electrode technology
    • Methodology: End-point, Fixed-time, Kinetic, optional ISE Single/Dual reagent chemistries, monochromatic/bichromatic Linear/non-linear multi-point calibration
    • Programming: Open system with user defined profiles and calculation chemistries Original system pack reagents ready to use
    • Reagent/Sample Handling: Reagent/Sample tray: 40 positions for reagents and 40 positions for samples in cooling compartment (2~12˚C)

Optical System:

    • Light Source: Halogen-tungsten lamp
    • Photometer: Grating system, reversed optics
    • Wavelength: 12 wavelengths, 340nm, 380nm, 412nm, 450nm, 505nm, 546nm, 570nm, 605nm, 660nm, 700nm, 740nm and 800nm
    • Absorption range: 0~3.3Abs (10mm conversion)
    • Resolution: 0.0001Abs

Reaction System:

    • Reaction rotor: Rotating tray, containing 80 cuvettes
    • Cuvette: Reusable, optical length 5mm
    • Reaction volume: 150~500μl
    • Reaction temperature: 37˚C
    • Temperature fluctuation: ±0.1˚C
    • Mixing System: Independent mixing bar

Reagent volume:

    • R1: 10~350μl, step by 1 μl
    • R2: 10~200μl, step by 1 μl
    • Sample volume: 2~45μl, step by 0.1 μl
    • Reagent/Sample probe: Liquid level detection and tracing, vertical & horizontal collision protection and inventory checking
    • Probe cleaning: Automatic washing for both interior and exterior Carry-over < 0.1%
    • Automatic sample dilution: Pre-dilution and post-dilution Dilution ratio up to 1: 200
    • Internal Bar Code Reader (optional): Used for sample and reagent programming Applicable to various bar code systems of Codabar, ITF (Interleaved Two of Five), code128, code39, UPC/EAN, Code93 Capable to communicate with LIS in bi-directional mode
    • Cuvette Washing: 8-step washing station with pre-warmed detergent and water

Control and Calibration:

    • Calibration mode: Linear (one-point, two-point and multi-point), Logit-Log 4P, Logit-Log 5P, Spline, Exponential, Polynomial, Parabola
    • Control software: Westgard multi-rule, Cumulative sum check, Twin plot

Operation Unit:

    • Operation system: Windows® XP Professional/Home SP2 or above Windows® 7
    • Interface: RS-232

Working Conditions:

  • Power Supply: AC 200~240V, 50/60Hz, ≤1500VA or AC 100~130V, 50/60Hz, ≤1500VA
  • Temperature: 15-30˚C for operation
  • Humidity: 35-85% RH
  • Water consumption: ≤4.5L/hour
  • Dimension: 860mm (W) x700mm (D) x625mm (H)
  • Weight: 130 Kg

BS-390: High End Fully automated Biochemistry Analyzer With Laundry

  • Fully automated Biochemistry Analyzer with Discrete, random access, fully automated, Constant 300 tests per hour with 75 sample positions.
  • Automatic probe cleaning, liquid level detection and collision protection mechanism.
  • Comes with 8 step laundry.”
  • Discrete, random access, fully automated.
  • Constant 300 tests per hour, up to 450 tests per hour with ISE.
  • Onboard capacity of 75 sample positions, up to 58 chemistries plus 3 ions.
  • Refrigerated reagent compartment.
  • Automatic probe cleaning, liquid level detection, vertical and horizontal collision protection.
  • 8-step auto washing system.
  • 12 wavelengths: 340~800nm.
  • Automatic dilution for abnormal sample.
  • Internal bar code reader (optional).
  • Bi-directional LIS interface.
    • System Function: Fully automated, Discrete, Random access for routine, STAT, Urine and homogeneous immuno assays; STAT sample priority
    • Reagent probe: Liquid level detection, vertical and horizontal collision protection
    • Probe cleaning: Interior and exterior automatic washing Carry-over< 0.1%
    • Throughput: 300 tests / hour, up to 450 tests / hour with ISE Measuring principles: Absorbance photometry, Turbidimetry
    • Methodology: End-point, Fixed-time, Kinetic, optional ISE Single / Dual reagent chemistries, monochromatic / bichromatic Linear / non-linear calibration
    • Programming: User defined profiles and calculation

Sample Handling:

    • Sample tray: 75 positions for primary or secondary tubes and sample cups
    • Sample volume: 2~45 l, step by 0.1 l
    • Sample probe: Liquid level detection, vertical and horizontal collision protection
    • Probe cleaning: Interior and exterior automatic washing
    • Carry-over< 0.1% Automatic sample dilution:
    • Pre-dilution and post-dilution ratio up to 1:150
    • Dilution vessel: Cuvette
    • Internal Bar Code Reader (optional): Used for sample and reagent programming; applicable to various bar code systems including Codabar、ITF、Code128、 Code 39、UPC / EAN、Code 93. Bi-directional LIS interface
    • ISE Module (optional):
    • Optional selection of K+, Na+, Cl− Throughput: Up to 225 tests per hour

Reagent Handling:

    • Reagent tray: 60 positions in refrigerated compartment (2~8˚C)
    • Reagent volume: R1: 150~350 l, step by 1 l; R2: 20~200 l, step by 1 l;

Reaction System:

    • Reaction rotor: Rotating tray, 72 cuvettes with automatic washing
    • Cuvette: Optical length of 5mm
    • Reaction volume: 150~360 l Operating temperature: 37˚C Temperature fluctuation: ±0.1˚C
    • Mixing system: Integrated mixers system

Optical System:

    • Light Source: Halogen-tungsten lamp
    • Photometer: Reversed optics, grating photometry
    • Wavelength: 340nm、380nm、412nm、450nm、505nm、 546nm、570nm、605nm、660nm、700nm、 740nm、800nm
    • Absorbance range: 0~3Abs Resolution: 0.001Abs

Control and Calibration:

    • Calibration mode: Linear (one-point, two-point and multi-point), Logit-Log 4P, Logit-Log 5P, Spline, Exponential 5P, Polynomial 5P, Parabola
    • Control rules: Westgard multi-rule, Cumulative sum check, Twin plot

Operation Unit:

    • Operation system: Windows® XP Professional/Home SP2 or above Windows® Vista Home/Business
    • Interface: RS-232

Working Conditions:

  • Power Supply: AC 100~240V, 50 / 60Hz, 1000VA Temperature: 15~30˚C
  • Humidity: 35~85% Water consumption: 10L/hour
  • Dimension: 990mm x 693mm x 1135mm (WxDxH) Weigh: 200 Kg
biochemistry_analyzer_bs390_big
BS-480

BS-480: High End Fully automated Biochemistry Analyzer With FDA Approval

Fully automated Biochemistry Analyzer with Discrete, random access, fully automated, Constant 400 tests per hour upto 560 with ISE. with 90 sample positions. Automatic probe cleaning, liquid level detection and collision protection mechanism. Comes with 8 step laundry.

  • Discrete, random access, fully automated
  • Constant 400 tests per hour, up to 560 tests per hour with ISE
  • Upto 80 on-board chemistries & 3 ions.
  • Refrigerated reagent compartment at 2~10 °C.
  • 90 sample positions and 90 cuvettes.
  • Automatic probe cleaning, liquid level detection & collision protection.
  • 8-step auto washing system.
  • 12 wavelengths: 340~800nm.
  • Minimum reaction volume 120 μl.
  • Intelligent auto-recovery from probe collision.
  • Advanced multi-functional and easy-to-use software.
  • Enhanced liquid system with easy maintenance.
  • Dedicated system pack reagents.
    • Sample tray: 90 positions for primary or secondary tubes and sample cups
    • Sample Volume: 1.5~45µl, step by 0.1µl
    • Sample probe: Liquid level detection, clot detection and collison protection
    • Probe cleaning: Interior and exterior automatic probe washing
    • Diluton vessel: Cuvette
    • Internal bar code reader (optional)
    • ISE Module (optional)
    • Optional selection of K⁺Na⁺Cl⁺
    • Throughput: Up to 240 tests per hour
    • Reagent Handling
    • Reagent tray: 80 positions in refrigerated compartment (2~10°C)
    • Reagent volume: 10~350µl
    • Reaction System:
    • Reaction rotor: Rotating tray, 90 cuvette with automtic washing
    • Cuvette: Optical length 5mm
    • Reaction volume: 120~360µl
    • Mixing system: 2 independent mixers
    • Optical System:
    • Light source : Halogen-tungsten lamp
    • Wavelength: 340nm, 380nm, 412nm, 450nm, 505nm, 546, 570nm, 605nm, 660nm, 700nm, 740nm, 800nm

Control and Calibration :

  • Control Rules: Westgard multi-rule, Levy-Jennings, Cumulative sum check, twin plot

BS-800: Modular, Fully automated Biochemistry Analyzer

BS-800 Modular System combines innovation and high performance into an integrated solution.with an expandable platform,an ever more complete line of chemistry reagents, as well as calibrators and controls and is accurate, convenient & cost effective.

  • Constant 800 photometric tests per hour, up to 600 tests per hour for ISE (K, Na, Cl).
  • 24-hour refrigeration for reagent tray and inner sample tray.
  • Permanent cuvettes with 8-stage auto-washing station
  • Two 3-head mixing bar units.
  • Minimum 100μl reaction volume.
  • Reversed grating system with 12 wavelengths.
  • Bubble detection for reagent and sample probes.
  • Support up to 4000m altitude application (external air pump optional).
  • Bi-directional LIS interface.
    Sample Handling:
  • Sample tray: 140 positions including 25 cooling positions for calibrators and controls
  • SDM: 300 samples capacity with 30 racks
  • Sample volume : 1.5~35 µL, step at 0.1µL
  • Sample probe: Liquid level detection, clot detection and collision protection

Reagent Handling:

  • Reagent tray: 120 positions in coaxial disks
  • Reagent volume: 15~300 µL, step at 0.5 µL
  • Reagent probe: Liquid level detection, bubble detection and collision protection

External Bar Code Reader (optional): Reaction System:

  • Reaction volume: 100~360 µL

Optical System:

  • Light source: Halogen tungsten lamp
  • Photometer: Reversed optics, grating photometry
  • Wavelength: 340nm, 380nm, 412nm, 450nm, 505nm, 546nm, 570nm, 605nm, 660nm, 700nm, 740nm, 800nm

ISE Module (Optional):

  • Principle: Indirect K+, Na+ Cl-, with 22 µL sample aspiration

Control and Calibration:

  • Control rules: Westgard multi-rule, Twin plot
BS-800

Hematology

Triscanduo

Triscan Duo: Advanced 3 Part Hematology Analyzer VET Model Available

Double Chamber ׀ Two Reagents ׀ Barcoded Triscan Duo is the Latest model in Hematology Analyzers combining Reliable Hardware with intuitive Software. Designed for Easy Operations and Convinient Data Communications/Interface. An Affordable Product without compramise on Performance

  • Type: Double Chamber
  • Measuring time: 60sec/ Sample, 60T/ Hr
  • Parameters : 20 parameters + 3 Histograms
  • Memory: 100,000
  • Sample Volume: 10ul Whole Blood
  • Reagents: 2 Reagents
  • Touch Screen: 8 inch intuitivetouch screen with Internal Thermal Printer
  • Measuring Mode: Venous,prediluted and capillary mode
  • Barcode : External Barcode for reagent & samples
  • Support LIS and external printer
  • Net weight: 19 Kg
  • CE marked
    Principle
  • Impedance for cell counting
  • Cyanide-free method for HGB

Parameters

  • WBC, Neu#, Lym#, Mid#, Neu%,Lym%, Mid%, RBC, HGB, HCT, MCV, MCH, MCHC,RDW-SD, RDW-CV, PLT, MPV, POW, PCT, P-LCR

Histograms

  • WBC, RBC, PLT histogram

Languages

  • English
  • Quality control: 3 Level QC, LJ Graph.
  • emperature:18°C-35°C
  • Power:AC 100-240V, 50/60±1 Hz

Dimension

  • 410mmx43Smmx472m

Sample volume

  • Venous mode : 10IJL Capillary
  • mode : 10uL
  • Prediluted mode: 20uL

BC5300: 5 Part Hematology Analyzer with Flow Cytometry and Impedence Technologies Combined

The new BC-5300 Auto Hematology Analyzer is a bench top system, combining three mainstream technologies: laser scatter, flow cytometry and chemical dye to provide reliable and accurate 5-part differentiation on white blood cells.

Tailored for independent labs, BC-5300 is compact, economical and very friendly to use. Only 20uL of blood is enough to have a 27-parameter report within 1 minute.

White blood cell differentiation
The semi-conductor laser system collects forward scatter and side scatter information of white blood cells in a flow cell and differentiates lymphocytes, monocytes, neutrophils and eosinophils according to the cell size and granule complexity.

Specific chemical dye treatment to eosinophils can separate from neutrophils in DIFF scattergram. In addition, basophils are measured in an impedance channel upon lysing action on the RBC and other WBC groups.

Advanced impedance measurement for RBC and PLT Mindray patented digital sheath flow and weak signal detection techniques are applied to ensure accurate counting and sizing of RBC and PLT in the impedance channel.

Dynamic discriminator adjustment can improve the separation of RBC and PLT groups when large platelet and micro RBC are present.

Benefiting from these improvements, MCV, RDW and PLT performance are greatly enhanced.

Flag abnormity
Besides of the routine CBC+DIFF results, BC-5300 offers data result for large immature cell and atypical lymphocytes in infectious diseases. Intelligent flagging system can help doctors fish out suspicious cases for further validation check.

For example, microcytosis, anemia, eosinophilia, PLT clump and so on.

The manual microscopic check results and morphology description can be print together in the final report.

    Principles
  • Flow Cytometry (FCM), Semi-conductor Laser scatter, chemical dye method, independent Basophil channel Impedance method for WBC, RBC, PLT counting Cyanide-free reagent for Hemoglobin test

Parameters

  • 27: WBC, Lym%, Mon%, Neu%, Eos%, Bas%, Lym#, Mon#, Neu#, Eos#, Bas#, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, LIC%, LIC#, ALY%, ALY# 3 histograms and 1 scattergram

Throughput

  • Up to 60 samples per hour

Sample Volume

  • Whole Blood: 20 u Lµ
  • Prediluted: 20 uL

Test Mode

  • CBC CBC+DIFF

Performance

  • Data Storage Capacity Up to 40,000 results including numeric and graphical information

Communication

  • LAN Port supports HL7 protocol
  • Operating Environment Temperature: 15°C~30°C Humidity: 30~85%
  • Air Pressure: 70~106 kPa

Power Requirement

  • A.C 100-240V ≤ 300VA 50/60Hz

Dimension and Weight

  • 410mm(L) x 470mm(W) x 530mm(H)
  • Weight: ≤45 Kg
BC-5300
BC-5300VET

BC5300VET: Versatile VET 5 Part Hematology Analyzer

BC-5300Vet Auto Hematology Analyzer pampers your need to get a reliable and accurate 5-part di¬erentiation hematology report for dog, cat, horse, rat, mouse, rabbit, pig, monkey, Goat, Camel.

With the state-of-art laser scatter, flow cytometry and chemical dye method, microsampling technology, friendly PC software and economical reagent kit, it is unprecedentedly easy and reliable to work with BC-5300Vet.

Principles Flow Cytometry (FCM), Semi-conductor Laser scatter, chemical dye method and independent Basophil channel Advanced DC impedance method for RBC and PLT counting Cyanide free reagent for hemoglobin test

    Microsampling
  • Only 20ul of blood (both whole blood mode and prediluted mode) is enough to start the CBC+DIFF test on BC-5300Vet. By single clicking, you can receive the complete report in 1min.

V-53 Reagent Kit

  • An innovatively designed reagent kit to sustain one month use (3 samples per day). The time-pack design ensures you lower leftover and easy replacement.

Laser Differential

  • The compact laser system collects forward and side scatter lights to differentiate cells based on size and granule complexity. The differential delivers you 23 parameters including the precise counting for lymphocytes, monocytes, neutrophils, eosinophils and basophils.
    Principles
  • Laser scatter, flow cytometry and chemical dye methods for WBC differential Advanced DC impedance method for RBC and PLT counting Cyanide free reagent for hemoglobin measurement

Parameters

 

  • WBC, Neu%, Lym%, Mon%, Eos%, Bas%, Neu#, Lym#, Mon#, Eos#, Bas#, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT

Sample Volume

  • CBC+DIFF 20 L
  • CBC 15 L
  • Whole blood and prediluted mode are supported

Throughput

  • Upto 60 samples per hour

Data Storage Capacity

  • Upto 40,000 results including numeric and graphic information

Communication

  • LAN port supports HL7 protocol
  • Operating Environment Temperature: 15°C~30°C Humidity: 30 % ~85 %
  • Air Pressure: 70 kPa~106 kPa

Power Requirement

  • A.C 100V-240V, ≤300 VA, 50/60 Hz
  • Performance
  • Precision
  • Linearity
  • Carryover
  • WBC ≤3.0% (6.00-15.00)×10⁹/L 0.00-99.99×10⁹/L ≤ 0.5 % RBC ≤2.0% (3.50-10.00)×10¹²/L 0.00-16.99×10¹²/L ≤ 0.5 % HGB ≤2.0% (110-180) g/L 0-250g/L ≤ 0.6 % PLT ≤6.0% (150-500)×10⁹/L 0-2999×10⁹/L ≤ 1.0 %

BC 5380: Most Popular 5 Part Hematology Analyzer with Autoloader

  • With autoloader and closed tube sampling functions, BC-5380 can bring you safety and convenience.
  • It’s design, powerful software, user friendly operation, high performance, cost efficiency and dynamic report format ensure end user satisfaction.
  • mpedance method, advanced flow cytometry, laser scatter light, chemical dye method, all these main-stream technologies are combined together in one package to assure accuracy and consistency of test results.
  • 5-part differentiation, 27 parameters, 1 scattergram and 3 histograms
    • 5-part differentiation, 27 parameters, 1 scattergram and 3 histograms
    • Semi-conductor Laser Scatter combined with Chemical Dye method, Advanced Flow Cytometry
    • Only 20ul sampling volume
    • Up to 60 samples per hour
    • 2 sampling modes: Autoloader and Closed Tube
    • 2 counting modes: Whole blood and

Prediluted

    • Capability to flag abnormal samples

Large storage capacity: up to 40,000 samples laser light scatter and chemical dye method

    • BC-5380 utilizes reliable and economical semi-conductor as light source for flow cytometry system with unique chem ical dye reagents treatm ent. Through analysis of different angle scatters, cell size, granularity and complexity, WBC 5-part differentiation can be accurately achieved.

Dual Acceleration Flow cell system

    • Dual acceleration flow cell system ensures more stable and single cell flow, which allow for accurate measurement via laser scatter.

Safe and Efficient

    • Autoloader with 30 samples capacity and random access to enhance walk away autom ation No need to un-cap sample tubes, Closed tube mode can safeguard from operational hazards Up to 60 tests per hour with autoloader Up to 50 tests per hour with Closed tube
    • Flexible and Convenient .~. .-= .””‘l-• Customizable, versatile and powerful operation software Troubleshooting with just one press of a button
    • No daily maintenance
    • STATsamples can be inserted randomly

Cost-effective and Accurate

  • Only 4 reagents for routine sampling including Diluent, LEOI Lyse, LEO II Lyse and LH Lyse Combination of advanced technologies enhancing accuracy
  • Comprehensive quality assurance
    Principles
  • Impedance method for WBC, RBC,PLT counting Cyanide-free reagent for Hernoglobin test Flow Cytometry (FCM), Semi-conductor Laser scatter, chemical dye method, independent Basophil channel

Parameters

  • WBC, Lyrn%, Mon%, l\Ieu%, Eos%, Bas%, Lym#, Mon#, I\leu#, Eos#, Bas#, RBC,HGB, HCT, MDI, MCH, MCHC, ROW-CV,ROW-SO,PLT, MPV, POW, PO, LlC%, LlC#, ALY%, ALY# 3 histoqrams and 1 scatter gram

Sampling Mode

  • Autoloader
  • Closed Tube

Throughput

  • Upto 60 samples per hour with autoloader
  • Upto 50 samples per hour with closed tube

Sample Volume

  • Whole Blood: 201J1
  • Prediluted: 201J1

Data Storage Capacity

  • upto 40,000 results including numeric and graphical information

Communication

  • LAI\I Port supports HL7 protocol

TestMode

  • CBe CBC+OIFF

Performance

  • Operating Environment
  • Temperature: 15 ‘C -30 ‘C
  • Humidity: 30-85%
  • Air Pressure: 70-106 kPa
  • Power Requirement L A.C100-240V ~300VA 50/60Hz

BC-5380
BC-6800

BC-6800: 7 Part Hematology Analyzer With SF Cube Technology

  • Fully Automatic Cell Counter.
  • Based on SF Cube cell analysis technology for WBC, 5-Part diff, NRBC, RET and PLT-O
  • Focusing Flow-DC method for RBCand PLT
  • 37 reportable parameters.
  • In 2D scattergram, the cell clusters appear close to each other. Whereas with the 3D scattergram, clusters are well separated which helps in detecting abnormal cells, if any.
  • SF Cube Cell Analysis Technology
  • Small Cube, Big Difference HFC%* MON% LYM% DIFF IMG% Left shift (flag) (NEU+BAS)%
  • DIFF scattergram of BC-6800 differentiates WBCs into 4 parts and also provides valuable parameters like HFC*, IMG* InR* and Abnormal Flags such as Left shift, NRBC, PLT Clump, Atypical Lymphocyte.
  • IMG(#, %) provide information about the presence of immature granulocytes, if any, including promyelocytes, myelocytes, metamyelocytes, immature eosinophils and immature basophils.
  • BC-6800 extends WBC diff to 6-part, on every sample, by including immature granulocyte.
  • NRBC (flag) PLT Clump (flag) InR‰* EOS%
  • HFC*(#, %) parameters alert the user of the presence of high fluorescent cell population, if any, such as blasts and atypical lymphocytes.
  • BASO Basophils are counted in a dedicated channel that detects information about cell volume and cellular complexity. This provides more accurate and reliable basophil results.
    • Principles: SF Cube cell analysis technology for WBC, 5-Part diff, NRBC, RET and PLT-O Focusing Flow-DC method for RBC and PLT Cyanide free hemoglobin measurement
    • Parameters: 37 reportable parameters (whole blood): WBC, Lym%, Mon%, Neu%, Bas%, Eos%, IMG%, Lym#, Mon#, Neu#, Eos#, Bas#, IMG#; RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, RET%, RET#, IRF, LFR, MFR, HFR, RHE, NRBC#, NRBC%; PLT, MPV, PDW, PCT, P-LCR, P-LCC, IPF
    • 17 research parameters (whole blood): HFC#, HFC%, WBC-R, WBC-D, WBC-B, WBC-N, RBC-O, PLT-O, PLT-I, PDW-SD, InR#, InR‰, Micro%, Micro#, Macro%, Macro#, MRV
    • 7 reportable parameters (body fluid): WBC-BF, TC-BF#, MN#, MN%, PMN#, PMN%, RBC-BF
    • 7 research parameters (body fluid): Eos-BF#, Eos-BF%, Neu-BF#, Neu-BF%, HF-BF#, HF-BF%, RBC-BF
    • 2 histograms for RBC and PLT
    • 3 scattergrams (3D) for DIFF, NRBC and RET
    • 6 scattergrams (2D) for DIFF, BASO, NRBC, RET, RET-EXT, PLT-O
    • Performance: Parameter Linearity Range Precision Carryover
    • WBC 0-500×10⁹/L ≤2.5% (≥4×10⁹/L) ≤1.0%
    • RBC 0-8×10¹²/L ≤1.5% (≥3.5×10¹²/L) ≤1.0%
    • HGB 0-250g/L ≤1.0% (110-180g/L) ≤1.0%
    • HCT 0-75% ≤1.5% (30%-50%) ≤1.0%
    • PLT 0-5000×10⁹/L ≤4.0% (≥100×10⁹/L) ≤1.0%
    • RET# 0-0.8×10¹²/L ≤15% (RBC≥3×10¹²/L; / 1%≤RET%≤4%)

Sample Volume:

    • Predilute mode (capillary blood), Open vial 40 L
    • Manual mode (whole blood), Open vial 150 L
    • Autoloader mode (whole blood), Closed vial 200 L
    • Manual mode (body fluid), Open vial 150 L

Throughput:

  • Upto 125 samples per hour (CBC+DIFF)
  • Upto 90 samples per hour (CBC+DIFF+RET)
  • Upto 40 samples per hour (body fluid)
  • Loading capacity: Up to 100 sample tubes
  • Mode: CBC, CBC+DIFF, CBC+RET, CBC+NRBC, CBC+DIFF+RET, CBC+DIFF+NRBC, CBC+DIFF+RET+NRBC, RET
  • Data storage capacity: Up to 100,000 patient results including all numeric and graphical information
  • Printout: Various printout formats and user-defined formats available
  • Operating environment Temperature: 15ºC~32ºC Humidity: 30%~85%

Immunology

Elisa Reader: 96 Wells Open System can accommodate wide range of your EIA testing needs

  • Touch screen, 7’8” large color LCD display
  • Opaque lid to prevent stray light
  • Built-in thermal thermal recorder, optional external printer
  • Memory: 20, 000 results
  • Multiple Curve-Fit options
  • Fully automatic, Monochromatic, bi-chromatic readings.
  • Advanced Elisa Reader with touch screen Operations.
  • Long lamp life with dormancy feature
  • Optional Data management software
  • Open System to accomodate wide variety of Test Menu
  • Accomodates 96/48T well formats
  • Detection method : Absorbance.
  • Microplate Types : 96/48-Well plate.
  • Reading Speed : 5 Sec Single wavelength, 12 Sec Double Wavelength.
  • Wavelenghth Range : 400 ~ 799nm.
  • Measurement System : 3-Channel optical system.
  • Light Source : Halogen-Tungsten.
  • Reading Range : 0.001 ~ 3.500 Abs.
  • Measurement Range : 0.001 ~ 2.500 Abs.
  • Photometric accuracy : 0 ~ 0.2 A 0.001 Abs.
  • Wavelenghth accuracy : ±2nm.
  • Repeatability : <1%.
  • Stability :
  • Linearity : 0~1.5A < 1.0%.
  • Filters : 405nm, 450nm, 492nm, 630nm.
  • Connection port : 1 X RS232, 2X USB.
  • Shaking : variable speed and time shaking modes.
  • Input : Touch Screen
  • Display : Large Color LCD Display.
  • Size : 437X332X174nm.
  • Weight : 8.5 Kg.
ElisaReader
ElisaWasher2

Elisa Washer: 96 Wells Open System EIA Washer.

  • Compatible with flat, U- or V- bottom plates
  • 2 wash modes with interchangeable manifolds for 8- and 12- well strips
  • Automatic bottom detection
  • Variable Shaking and soak time
  • Stores up to 100 user-defined protocols
  • Sweep function with aspiration
  • Fully Automatic and Programmable Versatile EIA Wash System with 100 user-defined programmable Protocols.
  • It has Automatic Liquid Monitoring System with waste Alarm.
  • It has 0-9 programmable wash cycles and is complete user friendly.
  • Wash Programs : upto 100.
  • Wash Cycles : 0 to 9.
  • Dispensing volume : 50 upto 300µl.
  • Bottom collosion protection : automatic.
  • Residual volume per well : < 2µl.
  • Soaking time : 0 to 300 seconds.
  • Shaking time : 0 to 300 seconds.
  • Multiple options : 1X8, 1X12.
  • Standard bottle set : 2X2l(detergent), 1x2l(Waste).
    1X2l (distilled water).
  • Power supply : AC 100-240V, 50/60 Hz.
  • Temperature : 5° C ~ 35° C.
  • Humidity : 15~85 %.
  • Dimension: 400nm X 330nm X 145nm (LXWXH).
  • Weight : 7.2 Kg.

Maglumi 1000: Automated CLIA Analyzer

Nano magnetic, micro-beads seperation technology with a comprehensive test menu of 200 parameters. Comes with built-in RFID reader

  • Test Menu: Tumor Markers
  • Thyroid
  • Fertility
  • TORCH
  • Hepatic Fibrosis
  • Infections
  • EBV
  • Cardiac
  • Glyco Metablism
  • Bone Metabolism
  • Immunoglobulin
  • Kidney function
  • Drug Monitoring
  • Prenatal screening
  • Anemia
  • GH
  • Cortisol
  • ACTH
  • DHEA-S.
  • Principle of luminescence: Non-Enzyme Type Flash Chemiluminescence
  • 120T throughput/Hour with first result in 17 Minutes.
  • Random, Batch and Stat Modes
  • 144 Sample Primary Tubes with continiuos loading LIS Compatible
  • 15 reagents onboard Chilled
  • RFID tag with reagent info and built-in master curve
  • 2-Point calibration to adj the master curve Clot Detection
  • Liquid Level Detection functions.
  • 135X64x87 CM, Weiging 148 KGS
  • Operating Temperatures: 15-30’C

magulmi_1000_big

Maglumi 600: Chemiluminescence Immunoassay.

It Comes with Non-enzyme involved flash chemiluminescence, long stability of reagents, ABEI label, stable in acid and alkaline solution.

  • Throughput: up to 180 tests/hour
  • 24 hours ready to use
  • Time to first result:17 minutes.
  • Clot detection
  • Liquid level detection for reagent
  • Optional ratio for sample dilution
  • Auto dilution for high concentration sample
  • Constant 37? incubation
  • 15-30? Operating Temperature

Principle of Luminescence

  • ABEI label, stable in acid and alkaline solution

Main Features

  • Throughput: up to 180 tests/hour
  • 24 hours ready to use
  • Time to first result:17 minutes

Modes of Operation

  • Random, Batch and STAT

Sample Loading

  • Up to 16 sample tubes
  • Continuous loading, STAT available
  • Analyzer auto numbered
  • LIS connection
  • Refrigerated sample area with independent power supply

Reagent Loading

  • 4 reagents on board
  • Continuous loading
  • RFID reading all info of reagents
  • Refrigerated reagent area

Reagent Features

  • Integrated kit, ready-to-use, no pretreatment required
  • Integrated kit including calibrators
  • Using superparamagnetic microbeads
  • RFID tag storing all info of reagent
  • RFID tag with built-in master curve
  • 2-point calibration to adjust master curve

Other Features

  • Clot detection
  • Liquid level detection for reagent
  • Optional ratio for sample dilution
  • Auto dilution for high concentration sample
  • Constant 37? incubation
  • 15-30? Operating Temperature

Operation System

  • Windows 7
  • 10-inch notebook

Interconnection

  • Connection to LIS via COM or network card
  • Direct operation on LIS system to test

Dimensions Weight

  • 88×56×50cm, 56kg</ul>

BA600 Semi Auto Urine Analyzer

  • Test Parameters: Glucose, Bilirubin, Ketone, Specific Gravity, Blood pH, Protein, Urobilinogen, Nitritr, Leukocyte, Vitamin C.
  • Test Method: Dry strip color-developing method
  • Test Principle: Reflective photometry with dual-wavelength.
  • Memory: saves and auto-recovers more than 2000 samples results
  • 5-inch touch screen
  • 11 testing parameters
  • 100 samples/hour throughput
  • Dual wavelength reflectance test method
  • Auto & manual test mode
  • Thermal printer, auto & manual print
  • Support LIS and barcode
  • Less and easy maintenance
  • Principle: Dual wavelength reflectance method
  • Wavelength: 520nm, 640nm, 660nm
  • Parameters: Glucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Leukocyte, Vitamin C
  • Throughput: 100 samples/hour
  • Sample type: Urine
  • Test mode: Auto, manual
  • Waiting time: 15-40s
  • Display: 5-inch touch screen Memory: 2,000 results storage Language: English
  • Strip: Original GENIUS strip, 100 strips per bottle
  • Report format: Semi-quantitative and qualitative
  • Coincidence rate: ≥90%
  • Power supply: AC 100-240V 50/60Hz Working temperature: 10-30°C Atmospheric pressure: 86.0kPa-106.0kPa
  • Relative humidity: ≤70%
  • Size: 360mm×225mm×160mm
  • Net weight: 2.5kg

Urine Analysis

BA600
BA680

BA680 : Singel Strip HQ Urine Analyzer

    Easy operation:
  • User-friendly software allows you a better experience on the analyzer Intelligent sensor make urine strip detected automatically

Less maintenance:

  • Auto waste collection and alert make it less contaminated as well as less maintenance Power cut off protection make it even less break-down

High efficiency:

  • High throughput with 500 samples/hour, ideal for extensive sample laboratory Auto print result after the test is finished
  • Throughput: 500 samples/hour
  • Principle: Dual wavelength reflectance photometry
  • Touch screen
  • Sensor for auto detecting urine strip
  • Easy maintenance design with auto waste collection
  • Support auto&manual print
  • Original urine strips
  • Parameters: Glucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Leukocyte, Vitamin C
  • Input: 5.6 inch touch screen
  • Output: Internal thermal printer
  • Display: LCD
  • Lamp source: LED, 470nm, 525nm, 625nm
  • Storage: Up to 2,000 results Working Temperature:10-30°C Humidity: 20%-80% Dimension: 390*320*210mm
  • Power: AC100~240V,50/60 Hz,45W
  • Net Weight :6 kg

GE 200: Electrolyte Analyzer(Na,K,Cl)..

GE 200 , Automated Electrolyte Analyzer based on ISE(Ion Selective Electrode) principle having touch screen & requires low reagent volume.

  • Throughput: 60 samples/hour.
  • Principle: ISE(Ion Selective Electrode).
  • 5 inch touch screen.
  • Auto sampling, washing, calibration.
  • Low reagent consumption.
  • Maintenance –free electrodes, 12 months lifespan.
  • Auto-print and manual print.
  • CE marked.

 

    • Sample Type: Serum, CSF, etc.
    • Sample Volume:100-200uL.
    • Reagent Consumption: 2 times sample volume
    • Calibration: Automatic or On-demand Printer: Internal thermal printer Display:5 inch touch screen.
    • Storage: Up to 2,000 results.
    • Interface: RS-232, support LIS.
    • Working Temperature:15-30°C.
    • Humidity:20%-80%.
    • Dimension: 385*240*380mm

Reagents:

  • ISE Reagent Kit DS-I (NO.2) (A B clean) (550 mL).
  • Cleaning Solution (250 mL).
  • PVC Electrode Activation Solution (250 mL) Filling Solution (10 mL).
  • Ref. Filling Solution (10 mL) Deprotein Solution
  • (R1: Diluent 18mL/bottle*1; R2: Protease powder 100mg/bottle*6)
  • Power: AC100~240V,50/60 Hz,80W
  • Net Weight :6.7 kg