Fully Automated Blood Coagulation Analyzer :Operon XL-3200 is a true walkaway system,combining many technologies in one with 10 independent measuring channel,continuous automatic loading and discard of cuvettes.

Features:

  • Test Channels: Unique 10 measuring channels with Auto Dilution / duplicate testing options.
  • Methodology: By Viscosity (Magnetic induction) for clotting, Optical chromogenic, Optical immunoturbidimetric, Optical Delta Measurements.
  • Throughput: 180T/hour
  • Probe: Dual probe dedicated for primary and secondary reagents.
  • Cuvettes: 1000 cuvettes walkway transported & discarded by a dedicated sheathed manipulator.
  • Barcode: Integrated barcode for samples.
  • Connectivity: LIS compatible, bidirectional.
  • Sample: Can accommodate primary tubes/microtainers/transport tubes.
  • Reagent loading: Unique tilt design for minimal reagent wastage. Drawer type system for loading and unloading of reagents.

Technical Specification

Test Speed: 180T/Hour

Tests Supported: PT/APTT/Lupus/Protein C/Protein S/ APCR/Factor Assays/Fibrinogen by Viscosity Clotting Method (Magnetic Induction) D-Dimer by Immunoturbidmetry Method (with curves) AT-3/tPA/Protein C/S/Vwf etc by Chromogenic Method (with curves) PT Derived Fibrinogen by Delta Measurement Method.

Four directions ventilation and cooling function.

Sample Positions: 35 Positions for Samples with Integrated barcode reader.

Reagents: 20 Reagent position chilled at 15’C including 2 Dedicated Stirring Positions for PT and APTT.

X1 Dedicated Cleanser position (X1) for thorough cleaning of the probe tip both externally and internally. X2 Dedicated Washing Station on board that’s connected to a Detergent (X2)

Alarm: Insufficient Cuvettes, waste and reagent alarms, abnormal test results alarm.

Minimum Reaction Volume : 150μl

Minimum Sample Volume : 2.5μl

Minimum Reagent Volume :10 μl

Accuracy & Precision:

 

 

Repeatability

 

Parameter

Normal Plasma

Abnormal Plasma

PT(s)

3.0%

8.0%

APTT(s)

4.0%

8.0%

FIB(g/L)

8.0%

15.0%

TT(s)

10.0%

15.0%

D-Dimer(µg/ml)

15.0% ( 0-1µg/mL)

10.0% (≥5µg/mL)

ATIII (%)

5.0%

10.0%

Power Supply: 220V±22V

Temperature: 15°C~30°C

Relative Humidity: ≤70%

Atmospheric Pressure: 86.0kPa ~106.0kPa

Weight: 120Kg